Skip to main content

Configuration

Main​

The Main tab contains settings that affect all services.

General​

image image
  • Logs level: A way to categorize messages logged by the system. Each level represents a different message type, from general information to warnings and critical errors. These levels are useful to filter and manage logged messages.

    • Information: Messages providing general information about the application's status or behavior. They are not errors, but useful for tracking and analysis.

    • Debug: Detailed messages intended for developers to debug or diagnose issues during development or testing. These are very detailed and useful for understanding the application's internal flow.

    • Error: Messages indicating an error occurred during application execution.

    Logs are saved in the "logs" folder inside the root directory. There are three log types: "main", "dicom" (protocol trace) and "worklist". Logs can be accessed from the monitoring page or directly from the corresponding folder. Old logs are compressed to avoid excessive disk usage.

  • Language: Selects the system's default language, which is the one used on the login screen. Unlike the per-user language, this option has no Default entry, since it defines the system default itself.

  • Time zone: Selects the system's default time zone. The list includes the Default option and a search box (Search timezone…) to filter the available zones.

  • Download IDs: Generates a JSON file with software and hardware identifiers to generate the system license.

  • Upload License: Allows saving a license to activate the system. The license is activated after restarting the PACS.

  • Restart PACS: Restarts the PACS service.

Mail Server​

Defines the outgoing mail server used by the system to send email notifications. It is currently used by the Viewer feedback feature of the Proteus Viewer, which sends comments to the address configured in that tab.

  • Enabled: Turns email sending on or off. When disabled, any send attempt fails with the mail_server_disabled error. It is enabled by default.

  • Provider: Sending mechanism. It can be SMTP (an own or third-party mail server) or AWS SES (Amazon's mail service). Switching provider resets the fields to their default values, so they must be filled in again before saving.

SMTP​

  • SMTP Host: Host name or IP address of the SMTP server (for example, smtp.gmail.com). It is mandatory: if left empty, sending fails with the smtp_not_configured error.

  • Port: SMTP server port. The default value is 587, the usual one for STARTTLS.

  • STARTTLS: Enabled by default. Opens the connection and then encrypts it with STARTTLS. If disabled, the connection is made without encryption, which is only advisable for a relay inside the internal network.

  • Username and Password: SMTP authentication credentials. If both fields are left empty, the connection is made without authentication.

  • From Address: Address shown as the sender (From) of the emails, for example noreply@example.com. It is mandatory: if left empty, sending fails with the smtp_not_configured error.

AWS SES​

  • AWS Region: SES region where the sender domain or address is verified. The default value is us-east-1.

  • Access Key and Secret Key: Credentials of the IAM account used for sending. The account must have permission to send emails through SES.

  • From Address: Address shown as the sender (From) of the emails. It must be an identity verified in SES. It is mandatory: if left empty, sending fails with the ses_not_configured error.

DICOM​

The DICOM tab contains settings affecting all DICOM services.

image image

General​

  • AE-Title: The unique DICOM identifier. Default is "PROTEUS". Must not exceed 16 characters; uppercase only is recommended.

  • Max PDU data length: Maximum allowed length for Protocol Data Unit (PDU) data in the DICOM standard. In DICOM, data is sent in PDUs, which are data units used for communication between medical devices. The max PDU data length is the maximum amount of data that can be included in a PDU.

    It is an important parameter in DICOM systems configuration, as it affects communication efficiency and reliability between devices. If the max PDU data length is too small, data may need to be split across multiple PDUs, which increases complexity and processing time. On the other hand, if it is too large, communication errors or network congestion risks may increase.

    warning

    It is recommended to modify this parameter only when necessary, as it affects overall system performance.

  • Character set: Refers to the character set used to encode text values in DICOM data. These character sets specify how alphanumeric and special characters are represented in DICOM file text fields, such as patient names, study comments, and other metadata.

    Choosing the appropriate character set depends on application-specific requirements and the language used in medical data.

  • Stable age: The DICOM STORE-SCP service has no way of knowing when a study has finished receiving. The study is considered stable after a defined period; this does not guarantee at all that the study transfer has completed. The sending modality may have paused the transfer for a while and could resume in a few minutes. Although imprecise by design, this method is the best way to assume a study has been completely transferred. The default of 60 seconds is appropriate for most setups.

DICOM Clients​

By default, and to ease interoperability, the PACS can accept queries (C-FIND) and study reception (C-STORE) from unregistered clients. However, for security and to enable advanced features such as Query/Retrieve with C-MOVE, registering all DICOM clients is recommended.

When is registering a client mandatory?​

It is required to add a client to this list if:

  • You need to perform a Query/Retrieve using the C-MOVE service.
  • You want to send studies to that client from the PACS study browser.
  • The client requires a specific Transfer Syntax configuration.

Configuration for Query/Retrieve (C-MOVE)​

The C-MOVE service is the most common method to retrieve studies and requires bilateral configuration. Here is how to register a client to use C-MOVE.

Step 1: Add a new client

Click the "ADD" button to open the client configuration form.

dicom client modal dicom client modal

Step 2: Fill in client data

  • AE Title: The unique DICOM client name on the network (e.g., MY_VIEWER). Case-sensitive.
  • IP Address: The client device's IP address. The PACS will use this IP to send studies.
  • Port: The port the client is listening on for DICOM connections.

Once you save this configuration, the PACS can send studies to this client when a C-MOVE operation requests them.

info

What about C-GET? The C-GET service does not require this registration because image transfer happens on the same connection that initiated the query, so the PACS does not need to open a new connection to the client.

info

Configuration shortcut You can add a client to the list automatically by performing a DICOM ECHO from the client device to the PACS. The PACS will detect the new client and add it with its AE Title and IP address; you may need to edit it to add the correct port for C-MOVE.

Recent Activity​

The Recent Activity button, in the header of the clients table, opens a window listing the devices that stored images in the PACS, whether they are registered or not. It is the simplest way to discover the actual AE-Title and IP address of a modality before registering it as a client.

The table shows one row per AE-Title with:

  • AE-Title: Calling AE-Title used by the device when sending the images.
  • IP Address: IP address the last transfer from that AE-Title came from.
  • Last update: Date and time of the last recorded transfer.

If no device has sent images yet, the window shows No activity.

info

This record is runtime telemetry, independent of the configuration: it is accumulated during reception and flushed to the database every 30 seconds, so a transfer that just finished may take that long to appear. Records that stop being updated are kept for 90 days by default.

Storage​

The Storage tab contains settings for the containers that store DICOM medical image files.

General​

image image

The system supports multiple datastores, which can be file systems (local or shared folders) or via the S3 protocol.

  • Rotate on error: If a storage error occurs, the PACS automatically continues with the next datastore in order. If this option is enabled, when reaching the last datastore it returns to the first.

  • Checksum: Generates a numeric value of the stored file to verify data integrity and saves it in the database. The file can be validated each time it is accessed.

  • Replace Study Date/Time: If enabled, the system will try to obtain study date and time from other DICOM fields if the main fields (Study Date and Study Time) are missing or empty.

    • Study Date: Will be looked up in Series Date, Acquisition Date or Content Date.
    • Study Time: Will be looked up in Series Time, Acquisition Time or Content Time.

    If this option is disabled, the system will reject any study without a valid date (Study Date). For time (Study Time), if the field exists but is empty, the system assigns the default "000000.00". If the time field is completely absent, the study is also rejected.

  • Ingest Transcoding: Allows changing pixel encoding to reduce DICOM file size. By default, it only converts the file to JPEG2000 Lossless if the file does not have compressed encoding. To change all files regardless of compression, use the options labeled 'forced'.

  • Selected Datastore: The active datastore.

Gateway​

image image

When the PACS is enabled in gateway mode, the system will automatically forward all stored studies that have not been previously sent, in addition to new studies received.

The system supports configuring multiple gateways independently. Each gateway is configured in its own tab and may use a different sending method.

Sending can be performed via three methods:

  • STOW-RS: Uses the DICOM Web standard to send studies over HTTP. Requires the base URL and optionally authentication credentials.
  • STORE-SCU: Uses the traditional DICOM protocol (C-STORE). Requires selecting a previously registered DICOM client.
  • CLOUD: Sends studies to a Proteus Cloud instance. Requires an Access Key and a Secret Key provided by the cloud repository.

Workers​

The number of Workers determines how many sends can run in parallel.

  • If set to 1, the system sends studies one at a time, waiting for each to finish before starting the next.
  • Increasing the worker count lets the system send multiple studies (or instances) simultaneously, taking better advantage of bandwidth and processing large image volumes faster.

Manual Push​

If enabled, studies will not be sent automatically when received. Instead, the user must initiate sending manually from the study browser. Useful for finer control over which studies are sent to the gateway.

Caches​

Caches are temporary storage areas used to store frequently accessed data, in order to improve system performance. Each cache has its own section: DICOM Images Cache, DICOM Tags Cache and DICOM ZIPs Cache, in that order.

  • DICOM Images Cache: Stores uncompressed DICOM images, to avoid repeated compression and decompression of the same file.
  • DICOM Tags Cache: Stores DICOM tag values.
  • DICOM ZIPs Cache: Stores study ZIP files.
image image

All caches share the same configuration fields. Depending on whether it's a file system or S3, different options are available.

Caches have a maximum space. When they reach the upper size limit, the deletion process begins until the lower size limit is reached. Default values are recommended.

The reset cache button removes all its records. Cache folders can also be deleted manually. If the system can't find the file, it will recreate it without issues.

info

The system must be restarted for changes to take effect.

Autoremove​

image image

When PACS autoremove mode is enabled, all studies that entered the system before the configured number of days are deleted. Useful when using the PACS as a gateway and not storing studies long-term.

info

The system must be restarted for changes to take effect.

warning

Studies are not removed if the datastore is removed. The PACS will not be able to open the image because the container is missing.

Worklist​

This section configures the worklist.

General​

image image
  • Filter using calling AE Title: If enabled, the worklist returns to the client only studies whose 'Client AE Title' matches. Client AE Title corresponds to the JSON field 'scheduledProcedureStepSequence.scheduledStationAeTitle'.

  • Force Expiration: When loading a new study, the default expiration value is set, regardless of any prior expiration value.

  • Force Local Bucket: The worklist supports several buckets or virtual worklist servers. The LOCAL bucket is accessed via the system AE Title (configured in the Main section). Each bucket works independently. When this option is enabled, querying LOCAL returns all studies, whether or not they belong to the LOCAL bucket. Useful for connection tests.

  • Token Authentication: Allows using the endpoint (/api/v1/worklist/study) to upload studies without a token. For security, using the token is recommended.

  • Use MPPS: Removes the study loaded in the worklist via the DICOM MPPS service.

  • Split Names: Enables automatic splitting of full names into first and last name when loaded as a single string (without ^).

  • Expiration Timeout (Seconds): Default time used when forcing expiration.

  • Character set: DICOM character set configuration for the worklist FIND-SCP service.

Buckets​

image image

Studies are usually loaded into the LOCAL bucket. On query, if the option is enabled, the worklist filters the response by equipment. If you need to work with groups of equipment instead of individuals, you can create a new bucket or virtual server and assign an asterisk (*) to the client AE Title to match all AE Titles.

Scripts​

image image

There are cases where equipment needs more DICOM tags or needs them in a different way. If it is not possible to modify them on the RIS side, the worklist allows creating real-time filters that connect in series. Incoming JSON passes through all filters and is then stored in the database with the modifications. Scripts are programmed in the Rhai language.

The incoming JSON document variable is named 'src'. It has several methods to get or edit values. Every key has three possible data types: a value (string, int32, etc.), another document or a vector.

// Gets the value of the key. The value is returned as a String.
// src.get(key: String) -> String
let value = src.get("key");

// Gets the value of an element of the vector at the key. The value is returned as a String.
// src.get_vec(key: String, index: usize) -> String
let value = src.get_vec("key", 0);

// Gets the number of elements, only if the data is a vector.
// src.get_len(key: String) -> i32
let len = src.get_len("key");

// Gets the value of the key, filters only numeric characters and returns it as a String.
// src.get_numeric(key: String) -> String
let value = src.get_numeric("key");

// Gets the value of the key, filters only alphabetic characters and returns it as a String.
// src.get_alphabetic(key: String) -> String
let value = src.get_alphabetic("key");

// Edits the value of the key.
// src.set(key: String, value: String)
src.set("key", "new");

// Edits the value of the key for a vector.
// src.set(key: String, index: usize, value: String)
src.set("key", 0, "new");

As an example, this script modifies studies sent with the nuclear medicine modality.

// Read the modality from the incoming JSON document.
let mod = src.get("scheduledProcedureStepSequence.modality");

// Modify the document if modality is NM (nuclear medicine).
if mod=="NM" {
// Get the Accession Number value.
let acc_no = src.get("accessionNumber");

// Get the Accession Number value and filter only digits.
// For R123456 it becomes 123456.
let acc_no_numeric = src.get_numeric("accessionNumber");

// Get the AE-Title value.
let ae_title = src.get("scheduledProcedureStepSequence.scheduledStationAeTitle");

// Edit the Scheduled Procedure Step Start Time of the study to "000000.00"
src.set("scheduledProcedureStepSequence.scheduledProcedureStepStartTime", "000000.00");

// Edit two values with the Accession Number.
src.set("scheduledProcedureStepSequence.scheduledProcedureStepId", acc_no);
src.set("requestedProcedureId", acc_no);

// Set AE Title to * so the DICOM response contains
// the same AE Title of the equipment performing the query.
src.set("scheduledProcedureStepSequence.scheduledStationAeTitle", "*");

// Set the value with the numeric part of the Accession Number.
src.set("studyInstanceUid", "999999."+acc_no_numeric);

// Set the bucket value to the incoming AE Title.
src.set("bucket", ae_title);
}

// Return the JSON document with its modifications.
src

To add the script, click the ADD button. When you modify the script and save, the change is applied automatically. If a script does not compile, it is not used, but the others are. Use EVALUATE to validate that the script compiles correctly.

Report​

This section configures how DICOM Structured Reports (SR) are read.

When a device sends an SR to the PACS, the study is stored like any other, but its content — measurements, findings, patient data — is nested inside the DICOM document and cannot be used directly by an external system. The scripts in this section translate that content into a flat JSON, with the field names the HIS/RIS expects.

The read is requested through the /api/v1/report/execute endpoint, identifying the study by accessionNumber or by studyIuid. The PACS looks for the Comprehensive SR instance in the study, reads it, runs the matching script and returns the JSON the script produces.

info

Only the Comprehensive SR SOP Class is supported. A study with no SR instance, or with an SR of another type, returns an error.

Scripts​

image image

Each script is identified by a Name, which is only an internal label to recognize it in the list. The script is selected with the other three fields, matched exactly against the incoming SR:

  • Model: matched against the Manufacturer Model Name DICOM tag (0008,1090).

  • Manufacturer: matched against the Manufacturer DICOM tag (0008,0070).

  • Report: matched against the structured report title.

If no script matches that triple, the read returns an error. The Enabled option keeps the script loaded or out of service without deleting it.

Scripts are written in the Rhai language, like the Worklist ones. The incoming structured report variable is named 'src'.

DICOM header data is read with direct functions, all of which return a String:

// Patient.
src.patient_name(); src.patient_id(); src.issuer_of_patient_id();
src.patient_birth_date(); src.patient_birth_time(); src.patient_sex();
src.other_patient_names(); src.patient_comments();

// Study.
src.study_date(); src.study_time(); src.accession_number();
src.referring_physician_name(); src.study_description();
src.study_instance_uid(); src.study_id();

// Patient study.
src.patient_age(); src.patient_size(); src.patient_weight();
src.smoking_status(); src.additional_patient_history();
src.pregnancy_status(); src.last_menstrual_date();
src.reason_for_visit(); src.admission_id(); src.patient_state();

// Series.
src.series_date(); src.series_time(); src.modality();
src.series_description(); src.protocol_name();
src.series_instance_uid(); src.series_number();

// Equipment.
src.manufacturer(); src.institution_name(); src.institution_address();
src.station_name(); src.institutional_department_name();
src.manufacturer_model_name(); src.device_serial_number();
src.device_uid(); src.software_versions();

// SR document.
src.content_date(); src.content_time(); src.instance_number();
src.completion_flag(); src.completion_flag_description();
src.verification_flag(); src.preliminary_flag();

// SOP Common.
src.sop_class_uid(); src.sop_instance_uid();

The report content — the nested items — is read with the following functions. The returned items have four properties: 'key', 'value', 'units' and 'is_empty'.

// Returns the structured report title.
// src.get_report_title() -> String
let title = src.get_report_title();

// Gets an item from the patient section, by its name.
// src.get_patient(name: String) -> Content
let weight = src.get_patient("Patient Weight");

// An item missing from the report is not an error: it returns an empty item.
if !weight.is_empty {
print(weight.key); // "Patient Weight"
print(weight.value); // "70"
print(weight.units); // "kg"
}

// Gets every finding of an anatomic site with a given name.
// src.get_findings(finding_site: String, name: String) -> Findings
let findings = src.get_findings("Left Atrium", "Left Atrium Systolic Volume Index");

// The same finding may be repeated for different views or methods, so the
// right one is picked with a list of key/value pairs.
// subitems() -> SubItems
let items = subitems();
items.add("Image View", "Apical four chamber");

// Returns the first finding that contains all those pairs.
// findings.get_finding(subitems: SubItems) -> Content
let finding = findings.get_finding(items);

// Stops execution and returns the message as the read error.
// src.error(message: String)
if finding.is_empty {
src.error("Missing Left Atrium Systolic Volume Index.");
}

The value of the last expression in the script is the output: usually a map, which is the JSON the endpoint returns.

// Validate that the finding uses the expected units.
let values = [];

let weight = src.get_patient("Patient Weight");

if !weight.is_empty {
if weight.units != "kg" {
src.error("Invalid Patient Weight unit.");
}

values.push(#{
"name": "Patient Weight",
"value": parse_float(weight.value),
"unit": weight.units,
});
}

// Script output, which is the JSON response of the read.
#{
"subjectName": src.get_patient("Subject Name").value,
"accessionNumber": src.accession_number(),
"values": values,
}

To add the script, click the ADD button. When the change is saved the engine is rebuilt and the effect is applied automatically. Disabled scripts are not compiled.

warning

Unlike the Worklist scripts, a report script that does not compile stops the load: the scripts after it in the list are left unloaded. Use the EVALUATE button to validate that the script compiles correctly before saving.

Viewers​

DICOM WEB Services​

image image
  • Render endpoint optimization: When enabled, the render endpoint reads the original compressed file instead of transcoding it to Little Endian, as long as the transfer syntax is JPEG 2000 Lossless or JPEG Baseline (Process 1). For every other syntax, and if the direct decoder fails, the system automatically falls back to the transcoding path. Disabled by default.

The DICOM Web service uses a JWT token that differs from the user session token. This token can restrict study access and has a defined lifetime. To integrate with another viewer, generate a new token with its own expiration.

  • Duration: Lifetime of the token to generate. The options are 30 days, 3 months, 6 months, 1 year and 2 years.

  • CREATE TOKEN: Generates the token and copies it to the clipboard. The token is issued with save and delete permissions; delete is only granted when the Proteus Viewer Remove WADO option is enabled.

Proteus Viewer​

image image

The system ships with the Proteus viewer integrated. The viewer integrates natively with the system, but can also be used with other DICOM PACS that support the DICOM Web standard. Additionally, the viewer can store measurements, markers and perform DICOM prints using the system.

  • Enabled: Enables or disables the integrated viewer entirely.

  • Film Editor: Enables or disables the film editor module to organize images before printing.

  • Save WADO: Enables saving and persistence of the WADO-RS image cache to improve performance on subsequent loads.

  • Remove WADO: Enables the functionality to delete the WADO-RS image cache stored in the browser. It also gates the delete permission of the DICOM Web tokens that are issued.

  • Use local: Defines whether the viewer connects natively to the local PACS. The service can be used as viewer only with multiple image sources. Before disabling this option, configure the Sources properly. It is enabled by default, and disabling it paints the switch red because the viewer stops reading from the local PACS.

  • Save Print: Allows the viewer to save and persist the print jobs performed.

  • Use compatible TS: Forces the viewer to request transfer syntaxes natively compatible with web browsers. Enabled by default.

  • Download rendered image for OP: Allows the viewer to download already-rendered images (e.g., JPEG) for certain operations, optimizing bandwidth and performance.

Film images by modality​

Determines, per modality, which images the viewer puts in the film when it opens. It is configured with two lists and the arrows that move the selected modalities from one to the other:

  • Show all images: The film includes every image of the study.
  • Show only KOs images: The film is limited to the study's Key Objects.

By default, every known modality (CT, MR, US, OCT, CR, DX, MG, XA, NM, PT, RF, SC, OT) sits in Show only KOs images. Modalities already stored in the configuration that are not in that list also appear in the selector.

Images per film page​

Maximum number of images shown at once in the Proteus viewer film. The only accepted values are 1, 2, 4 and 6; the default is 6.

Viewer feedback​

  • Viewer feedback: When enabled, the Proteus Viewer can send user comments by email. Disabled by default.

  • Feedback destination email: Address those comments are sent to. Sending uses the Mail Server configured in the Main tab, which must therefore be enabled and configured. The field is only editable with the option above enabled.

Sources​

Sources are the image origins the viewer uses when Use local is disabled. With Use local enabled, the viewer ignores this table and uses a single local source pointing at the PACS itself.

Each source is added with the SOURCE button and has the following fields:

  • Name: Label used to identify the source. Required.
  • PACS URL: Base URL of the origin PACS. Required.
  • WADO RS: URL of that source's WADO-RS endpoint, from which the viewer downloads the images. Required.
  • S3 URL: Public URL of that source's S3 bucket. If left empty, the viewer performs no direct S3 downloads.
  • Is Proteus: Marks the source as another Proteus instance. It lets the PACS request tokens from it on the user's behalf.
  • Proteus API Token: Token the PACS authenticates with against that remote instance to request the viewer token.
info

The Proteus API Token is never returned in API responses: the system strips it from the configuration it hands to the viewer.

Viewers​

image image

External viewers can be added or removed from the study browser, provided they have web integration. The table shows the image, the full URL, the supported platforms and whether it is enabled.

Each viewer is configured with the following fields:

  • Enabled: Shows or hides the viewer in the study browser.
  • URL Scheme: Scheme of the link that opens the application, for example radiant://, osirix://, weasis:// or http://.
  • URL Body: Rest of the URL, where the study parameters are given.
  • URL Body Encoded: Percent-encodes the URL body before concatenating it with the scheme. Required for viewers that take the whole command line as a single parameter, such as Weasis.
  • Image Width: Width, in pixels, the logo is shown with in the study browser.
  • Image File: Image identifying the viewer. Required to save; it is stored embedded in the configuration.
  • Linux, MacOS and Windows: Operating systems the viewer is shown on. If the browser does not report the platform, the viewer is shown anyway.

The URL Body accepts the following variables, which the study browser replaces when opening the viewer:

VariableReplaced by
<STUDY_IUID>Study Instance UID of the selected study.
<AE_TITLE>System AE-Title.
<PACS_URL>Current PACS URL.
<TOKEN>DICOM Web token generated for that study.
info

The token is only requested when the URL Body contains <TOKEN>.

Three example viewers ship by default, all disabled: RadiAnt (Windows), Horos/OsiriX (MacOS) and Weasis (Linux, MacOS and Windows).

Integrity​

Configures the integration with Integrity, the external reporting system.

  • Integrity URL: URL of the Integrity API.
  • Integrity API Key: Authentication key for that API.
  • Integrity Report URL: URL the viewer opens to show the report. It accepts the {reportAccess} variable, replaced by the reportAccess URL param received by the Proteus viewer.
  • Show reports: Shows or hides the reports in the viewer. Enabled by default.

The print server service simulates DICOM printers. Once the image is received, it can be sent to one or more paper printers.

The tab has five sections, in this order: Buckets, Logos, Resize, Color Change and Films Cache.

Buckets​

Each virtual printer is called a bucket. The table lists the configured buckets with the AE Title, Action, DICOM Film Size, DPI and Logo Name columns. The BUCKET button adds a new one, and clicking a row opens the same form to edit or delete it.

The bucket form has three sections: DICOM, Template, and a third one whose title changes with the selected Action.

DICOM​

image image
  • AE Title: The bucket AE Title. To send a print, use the server IP, port 11112 (if the default port has not been changed) and the configured AE Title. It is required and cannot repeat across buckets; on a repeat the form shows Duplicated AE-Title! and refuses to save.
  • DICOM Film Size: The film size. Besides Client, the options are 8INX10IN, 8_5INX11IN, 10INX12IN, 10INX14IN, 11INX14IN, 11INX17IN, 14INX14IN, 14INX17IN, 24CMX24CM, 24CMX30CM, A4 and A3. With Client, the default, the client (medical device) chooses this option.
  • DICOM Border: The film border color, White or Black. With Client, the default, the client (medical device) chooses this option.
  • DICOM Empty Image: The empty space color between film images, White or Black. With Client, the default, the client (medical device) chooses this option.
  • DPI: Pixel density per inch of the film. The available values are 300, 600, 1200 and 1600; the default is 1200.
  • Action: The action executed when the film is received. There are three possible actions:
    • Printer (Windows): Prints the film, lossless, on one or more paper printers (Windows only).
    • Lossy Printer (Windows): Prints the film, lossy, on one or more paper printers (Windows only).
    • API Endpoint: Sends the received image to an API endpoint via POST.

Template​

image image

The template allows modifying margins and adding the logo on the printed page. Margins are configured in pixels, while logo height and padding are set as a percentage of page size, to keep similar size across all pages.

  • Margin Top (px), Margin Bottom (px), Margin Left (px) and Margin Right (px): Page margins. A new bucket starts with 50 px on the top margin and 0 px on the rest.
  • Logo Name: Logo to use, picked among those loaded in the Logos section. The — value leaves the page with no logo.
  • Logo Height (%) and Logo Padding (%): Logo height and padding, as a percentage of the page size.
image

Printer​

image image

In the Printer or Lossy Printer section, the destination paper printers are configured. When a bucket receives a DICOM film, it looks up in the Resize table the conversion from film size to paper size and replaces it accordingly. For example, if I receive an 8INx10IN film, it converts it into an A4 paper print. Then, depending on paper size, it can be sent to one or more printers.

Each row of the printers table is built with the following fields and added with the + button:

  • Input Paper Size: Paper size that triggers this row, the result of the Resize conversion.
  • Printer Name: Destination paper printer. The list comes from the server operating system and is refreshed every time the dropdown is opened.
  • Paper Size and Paper Orientation: Size and orientation the job is sent with. With Default the printer's own configuration is kept.
  • Quality: Only on Lossy Printer (Windows). Set with a slider between 40 and 100, in steps of 5. With the lossless action the table always shows 100.

All prints are sent to the Default printer, configured in the subsection of the same name with the same fields (printer name, size, orientation and quality). It is required: without it the form refuses to save. The table rows are extra printers used based on the input paper size.

info

Using the lossy printer is recommended; otherwise the image size may exceed what the paper printer accepts, considerably slowing the load and potentially causing print issues. Generally, print quality is lower than lossy print when using between 85% and 95%.

API Endpoint​

image image

The system supports using an external application's endpoint to send the images to print. Another system can be used for printing the films. Sending uses HTTP POST and adds the authentication header via username and password.

  • URL Path: URL of the endpoint receiving the image.
  • Username and Password: Credentials for the authentication header.

All three fields are required: if any is left empty, the form refuses to save.

Logos​

image image

In this section you can add images that will be used in template logo configuration. The table shows each logo's Name and Image.

The LOGO button asks for the name first and then opens the crop editor to load the image. Clicking an already loaded image reopens that editor to replace it.

Resize​

image image

This table comes preconfigured to find the best conversion option between input film size and output paper size. If the film size is not listed, the default option is used.

The default conversion is:

Film Size (Input)Paper Size (Output)
8INX10IN, 8_5INX11IN, 10INX12IN, 10INX14IN, A4A4
11INX14IN, 11INX17IN, 14INX14IN, 14INX17IN, 24CMX24CM, 24CMX30CM, A3A3

The Default field, in the section header, sets the paper size used when the received film size is not in the table. Its default value is A3. The FILM button adds a new conversion; the input dropdown only offers film sizes that are not mapped yet.

warning

Modify this table only if necessary. Changes here affect all buckets.

Color Change​

image image

In this section you can configure brightness and contrast for a specific device. The device must be configured as a DICOM node and you must choose which bucket it will use. Generally, this configuration is not needed since adjustments can be done from the medical equipment or paper printer.

The COLOR button picks the AE Title and the Bucket the adjustment applies to. The AE Title comes from the devices that sent films to the print server, and the bucket from the configured buckets.

Once the row is added, Brightness and Contrast are edited directly in the table. Both start at 0; contrast accepts decimals.

Films Cache​

image image

A temporary storage area used to keep films received by the buckets. Depending on whether it's a file system or S3, different options are available.

Caches have a maximum space. When they reach the upper size limit, the deletion process begins until the lower size limit is reached. Default values are recommended.

The reset cache button removes all its records. Cache folders can also be deleted manually. If the system can't find the file, it will recreate it without issues.

info

The system must be restarted for changes to take effect.

This section contains the Proteus Web Viewer print client module configuration. It allows creating films from the Web Viewer and sending them to an external DICOM film printer.

The tab has four sections, in this order: General, Printers, Templates and Films Cache.

General​

image image
  • Browser Printer: Enables using the web browser printer in addition to DICOM film printers. Enabled by default.

Printers​

List where you create, edit or delete all external DICOM printers. The table shows the Name, Target, Film Size, DPI and Tags columns. The PRINTER button adds a new one, and clicking a row opens the same form to edit or delete it.

  • Name: Identifies the printer. It is required, takes no spaces and cannot repeat; otherwise the form shows Required field, No spaces are allowed or Duplicated name! and refuses to save.

  • Tags: Free-form tags used to group printers. Type in the field and add with Enter or by leaving the field; the dropdown suggests tags already used in the system. Remove them with each tag's cross.

  • Target: Defines where the film is sent.

    • Remote: An external DICOM printer. Enables the AE Title, IP and Port fields, all three required. A new printer starts with IP 127.0.0.1 and port 104.
    • Local: A bucket of this same PACS's Print Server. Enables the Bucket field, which lists the configured buckets and is required.

The DICOM print parameters are set with the following fields:

FieldOptionsDefault
Medium TypePaper, Clear Film, Blue Film, Mammo Clear Film, Mammo Blue FilmBlue Film
PriorityLow, Medium, HighMedium
Film DestinationMagazine, ProcessorProcessor
Film OrientationPortrait, LandscapePortrait
Film Size8INX10IN, 8_5INX11IN, 10INX12IN, 10INX14IN, 11INX14IN, 11INX17IN, 14INX14IN, 14INX17IN, 24CMX24CM, 24CMX30CM, A4, A38INX10IN
Magnification TypeReplicate, Bilinear, Cubic, NoneBilinear
Border DensityWhite, BlackWhite
Empty Image DensityWhite, BlackWhite
DPI300, 600, 1200, 16001200
Pixels FormatColor, Grayscale8, Grayscale12Grayscale12

Note the pixel format configuration. The printer can print in color, 8-bit grayscale or 12-bit grayscale.

info

When 12-bit grayscale (Grayscale12) is configured, the Web Viewer can use 12 bits, only if all image captures are 12 bits or more; otherwise it will use 8 bits.

Templates​

image image

You can add, edit or delete the print templates that come by default. Print templates are configured using a JSON document.

The Template Name dropdown selects the template to edit and loads its JSON into the Source field. The buttons are:

  • ADD: Asks for a name and loads a new 1x1 template with no sub-blocks into the editor.
  • SAVE: Stores the editor's JSON in the template list. It is only enabled when validation passes.
  • DELETE: Removes the selected template.

While typing, the system validates the JSON automatically. If it is valid it shows OK and enables SAVE; if not, it shows the parse error and keeps the button disabled.

info

The template is saved by its name: if a template with that name already exists, it is replaced; otherwise it is appended. So changing the name in the editor and saving creates a new template instead of renaming the existing one.

Eight templates ship by default: 1x1, 2x2, 2x3, 3x4, 4x5, 4x6, 5x7 and 1x1+3x2.

The parent document has the following fields:

  • name: The name shown in the Web Viewer.
  • sizeRows: Number of rows.
  • sizeCols: Number of columns.
  • childs: A vector that allows sub-dividing blocks. A new document is added for each sub-block and its internal configuration.

The child object has the following fields:

  • posRow: The block position by row. (From row 1 up to the maximum number of rows determined by sizeRows)
  • posCol: The block position by column. (From column 1 up to the maximum number of columns determined by sizeCols)
  • sizeRows: Number of sub-block rows.
  • sizeCols: Number of sub-block columns.
  • childs: A vector that allows sub-dividing blocks. A new document is added for each sub-block and its internal configuration.
warning

Every field is required, both in the parent document and in each sub-block. If one is missing, validation fails and the template cannot be saved.

As an example, the 1x1+3x2 template:

image
{
// template name
"name": "1x1+3x2",

// number of rows
"sizeRows": 2,

// number of columns
"sizeCols": 1,

// sub-block configuration
"childs": [
{
// row position of the block to modify
"posRow": 2,

// column position of the block to modify
"posCol": 1,

// number of sub-block rows
"sizeRows": 2,

// number of sub-block columns
"sizeCols": 3,

"childs": []
}
]
}

Films Cache​

image image

A temporary storage area used to keep films received from the Web Viewer. Depending on whether it's a file system or S3, different options are available.

Caches have a maximum space. When they reach the upper size limit, the deletion process begins until the lower size limit is reached. Default values are recommended.

The reset cache button removes all its records. Cache folders can also be deleted manually. If the system can't find the file, it will recreate it without issues.

info

The system must be restarted for changes to take effect.

Triggers​

This tab configures two independent triggers (webhooks), in this order: Stable Study Trigger and Completed MPPS Trigger. Each one sends a JSON document via HTTP POST to its own URL when the matching event occurs.

Both panels have the same fields:

  • Enabled: Turns sending on or off. Disabled by default.
  • URL: Address receiving the POST.
  • Auth: Request authentication. It can be none (no authentication), basic — which enables the User and Password fields — or bearer, which enables the Token field.

Each panel header has two buttons:

  • TEST: Sends a test POST to the configured URL, with the configured authentication and an empty document as the body. The button turns green if the destination answers with a success code, and red if it fails or returns any other code.
  • RESET: Empties that trigger's pending event queue.
info

Events are queued in the database and delivered one at a time. If delivery fails — because the destination does not answer or returns a non-success code — the event stays in the queue and is retried every 30 seconds, indefinitely. While the destination is down the queue builds up, so when the service comes back every backlogged event is sent. The RESET button empties that queue, discarding the pending events.

info

If neither trigger is enabled, the delivery process idles and re-reads the configuration every 120 seconds. The system does not need to be restarted for changes to take effect.

Stable Study Trigger​

image image

Sends a JSON via HTTP POST with the data of studies that have become stable. The time during which a study is considered stable is set in the Stable age field. Medical equipment can send all images at the end of the study or during the procedure. In the second case, the PACS will send the same study several times, indicating that modifications have been made to it.

The document sent carries the study data:

FieldContent
studyDbId, patDbIdInternal study and patient identifiers.
studyIuid, studyId, accessionNoStudy Instance UID, Study ID and Accession Number.
studyDate, studyTimeStudy date and time.
studyDesc, modsInStudy, cuidsInStudyDescription, modalities and SOP Class UIDs present in the study.
studyInstCount, seriesCountInstance and series counts.
patName, patId, patIdIssuer, patBirthDate, patSexPatient data.
instName, deptName, stationName, aeTitleInstitution, department, station and source AE-Title.
refPhysician, perfPhysician, readPhysicianReferring, performing and reading physician.

Completed MPPS Trigger​

image image

Sends a JSON via HTTP POST with the data of complete studies received by the DICOM MPPS service.

The document mirrors the received MPPS content: keys are camelCase and map to the procedure step's DICOM tags, among them sopInstanceUid, modality, patientName, patientId, patientBirthDate, patientSex, studyId, perfStationAeTitle, perfStationName, perfLocation, perfProcStepStartDate and perfProcStepStartTime, plus the code and referenced-study sequences. Fields that arrive empty are omitted from the JSON.

Backup​

Backup keeps a second copy of every stored image, in a datastore separate from the one used in Storage. A background process walks the instances that have not been backed up yet and copies them one at a time: it reads the file from the source datastore, writes it to the backup datastore, and records that copy's label and path in the database.

Studies that already have a copy are marked with the green archive box icon in the study browser.

The tab has two sections: General and Datastores.

General​

  • Enabled: Turns backup on or off. Disabled by default.

  • Selected Datastore: Datastore the copies are written to, picked among those defined in Datastores. The dropdown stays disabled while none exists.

  • Mode: Defines when the copy runs.

    • Real Time: The process copies continuously, as images come in.
    • Periodic: The process copies only inside a time window. It enables the seven weekdays and the Start Time and End Time fields. Outside that window backup pauses and pending instances pile up until the next one.
info

If End Time is earlier than Start Time, the window crosses midnight. In that case the selected day is the day the window starts: a 22:00 to 06:00 window with Monday enabled copies from Monday 22:00 through Tuesday 06:00.

warning

The process does not run in any of these cases: backup is disabled, no datastore is selected, the selected label matches no datastore in the list, or the selected datastore is marked Offline. In the last three a warning is written to the logs and the system retries every 30 seconds.

info

If an instance's copy fails, it is retried on the next pass. After five failed attempts the instance is dropped from the queue for good and the log records Backup abandoned after 5 retries, along with the last error. The retry limit cannot be changed from the interface.

Datastores​

List of the datastores available for backup. The table shows the Order, Label, Type, Path / Bucket and Offline columns. The DATASTORE button adds a new one, and clicking a row opens the same form to edit or delete it. The Offline column can be toggled straight from the table.

  • Order: The datastore's order number. A new datastore takes the next free number.
  • Label: Datastore name. It is required and is the value picked in Selected Datastore.
  • Type: FileSystem for a file system (local or shared folder) or S3 for the S3 protocol.
  • Offline: Marks the datastore as unavailable.

Type FileSystem has a single field:

  • Path: Folder the copies are stored in. It is required and the proposed value is ./backup.

Type S3 has the following fields:

  • S3 Region: Bucket region. Besides the AWS regions, the list includes the custom option for an S3-compatible service.
  • S3 Custom Region and S3 Custom Endpoint: Only with the custom region. Both are required.
  • S3 Access Key and S3 Secret Key: Credentials to access the bucket.
  • S3 Bucket: Bucket name. It is required.
  • S3 Path Style: Uses path-style addresses (https://endpoint/bucket/object) instead of virtual-hosted ones (https://bucket.endpoint/object). Usually needed on S3-compatible services such as MinIO.
info

The Offline option has two different effects depending on where it is used. On this tab's selected datastore, it stops the backup process. On a datastore in the Storage tab, it marks that datastore as no longer reachable and makes the PACS serve the images from the backup copy. If an instance in an offline datastore has no backup copy, the read fails.